Did you know that Comet exists in a US version as well?
dataloggers, thermometers, hygrometers...
English Česky Africa China Español Français Magyar Polski Sweden USA

Ethernet Multilogger - universal datalogger with 4 inputs

Battery powered datalogger with 4 universal inputs for measurement of 0-10V voltage, 0-20mA current, two-state signal and for counting pulses.
Signals are connected to four detachable terminal blocks.
Multilogger can be installed permanently to Ethernet network or work as portable device.
Included is traceable calibration certificate in accordance with EN ISO/IEC17025.

The Multilogger is a battery operated datalogger capable of measuring and recording sensor and electrical values through its Inputs.
Datalogger can be connected to Ethernet. Then measured values can be viewed on a web browser, as well as stored and downloaded for later analysis.

 

 

  • code
    M1300
  • Warranty
    3 years

Technical data

Measured value

Current
Voltage
Two-state signal
Pulses
DC VOLTAGE
Measuring range 0 to 10 V
Accuracy ±10 mV
Resolution better than 100 µV

DC CURRENT

Measuring range 0 to 20 mA
Accuracy ±20 µA
Resolution better than 0,1 µA

BINARY SIGNAL

Input voltage for level “switched on”/ <0,4 V/> 2 V
Minimum/maximum input voltage 0 V/+30 V DC
I mpedance of the “switched on”/ < 10 kOhm/> 300 kOhm

COUNTER INPUT

Range of counter 24 bit (16 777 215) the counter overflow can be allowed
Maximum input frequency 200 Hz

GENERAL TECHNICAL DATA

Operating temperature -10 to +60 °C
Channels 4x binary, counter, voltage and current input
Memory 1 000 000 values (noncyclic record), 600 000 values (cyclic record)
Recording interval 1s to 24 hours
Display and alarm refresh each 10 s
Recording mode noncyclic - data logging stops after filling the memory
cyclic - after filling memory oldest data is overwritten by new
Available temperature units degrees Celsius, degrees Fahrenheit
Communication interface Ethernet, USB, RS232
Ethernet communication protocols WWW, SNMPv1, XML, DATALINK
Alarm protocols E-Mail
Power AC adapter 230Vac/5Vdc (neccessary for Ethernet interface)
Battery life 3 - 4 weeks
Protection class IP20
Dimensions (without connectors) 178 x 95 x 37mm, without attached cables
Weight (including batteries) approx. 380 g
Warranty 3 years

Features

Applications:

Aplikace - Zdravotnictví a laboratoře

Healtcare and laboratories - Monitoring of storage areas at temperatures up to -200 °C.

Aplikace - Sklady

Food industry - Monitoring of critical processes according to the HACCP standard with the possibility of prompt notification of unforeseen events that could lead to the degradation of the goods.

Aplikace - Průmysl

Industry and production - monitoring temperature and humidity in food, pharmaceutical, aerospace etc.

Aplikace - Sklady

Production facilities and warehouses - Monitoring of storage conditions and production processes in the temperature range -200 ° C to + 600 ° C. Sending data to COMET Database or SCADA third-party systems.

Aplikace - Serverovny a datová centra

Metrology - COMET sensors together with COMET radiationshields are the ideal basis for weather stations for long-term weather, climate and air monitoring.

Aplikace - Serverovny a datová centra

Data centers, server rooms - Temperature and humidity monitoring, data center monitoring, RACK stands, including flood status reports, opening / closing of doors (windows), motion, smoke. SNMP monitoring, remote alarms by email and Syslog.

Aplikace - Automatizace budov a HVAC

Building management - Our instruments which measure temperature, humidity, atmospheric pressure and CO2 are an integral part of these industry-leading solutions.

Features description:

● the features may vary depending on the device type
● up to 6 hardware inputs for measuring and recording, 16 virtual channels for temperature, humidity, dew point, atmospheric pressure, current 0 - 20 mA and voltage 0 - 10 V (-60 to 140 mV), two-state events, pulses, CO2
● memory for 1 million values
● indication of alarm conditions via e-mail, SMS, integrated acoustic siren LEDs, stwo-state output to control other devices
● analytical software for data processing and configuration - COMET Vision
● communication interfaces - USB, RS232 and Ethernet (SNMPv1 and XML)

 

Warranty 3 years Traceable calibration certificate COMET Database Data report Datalogger Handheld device Two-state inputs 2 line LCD Internal memory Ethernet
 
AlarmBattery powerRecheargeable batteryIP protection
 
 

Up to 4 physical Inputs - TERMINALS

Multilogger M1300 inputs
Removable terminal blocks allows connection of:
  • Sensors with voltage output (range 0-10 V; accuracy ±10 mV)
  • Sensors with current output (range 0-20 mA; accuracy ±20 µA)
  • Sevices with two-state output (potential-less contact or voltage signal 0-30V); may be applied to input 3 and 4
  • Devices with pulse signal (potential-less contact or voltage signal 0-30V); may be applied to input 4

Evaluation of up to 16 variables

Multilogger - LCD detail
Up to 16 variables may be calculated from the four connected probes / sensors. This is the sum of the measured and calculated values. The calculated values can be a further expression of moisture – dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy. It can be also the result of inter-channel conversions, e.g. the difference of two connected temperature probes.Multilogger LCD display detail.

The measured values are stored in the internal non-volatile memory. Memory capacity is up to 1 million values. Logging mode can be adjusted as non-cyclic, when logging stops after filling the memory, or cyclic, when after fulfilling the memory oldest stored values are overwritten by new ones.

Measured values are displayed on the screen with backlight option and the ability to change the size of display digits

Power supply

Multilogger - Power Supply
Multilogger - Power Supply

Multilogger can be powered by AC adapter or by rechargeable batteries with charging capability. Batteries are always a part of the device (you can also use standard alkaline AA batteries). Batteries power the device for several months.

While using the AC adapter, the batteries provide backup operations and provide recording and measurement during power failure. Thanks to the Ethernet output the device informs about the failure via e-mail message.

Alarm limits and alarm indication

Multilogger - Alarm limits

It is possible to set two independent alarm limits for each channel (i.e. measured or calculated value).

Alarm signalization can be acoustic (built-in beeper), optical, sending an e-mail alert or an alarm output. The alarm output is for connecting an external audio or visual alarm.

Communication interfaces

Multilogger - Alarm limits

Multilogger is always equipped with USB connector for communication, data analysis and also for charging of the rechargeable batteries.

Included Ethernet interface allows to:

  • send an e-mail when an alarm state starts or ends
  • use DATALINK: display current values or download values from memory to COMET Database in your PC. Software COMET Database contains many useful tools for data analysis - graphs, tables, statistics, etc.
  • simple wireless communication with Wi-Fi TP-LINK router
  • third-party applications to read the actual measured values using universal protocols SNMPv1 and XML
  • view current measured values using your web browser

Output ALARM OUT

Multilogger - Alarm limits and alarm indication
  • Output can be used for controling other devices when alarm is launched, such as external buzzer, telephone dialer etc.
  • Type of output - open collector transistor
  • Max. switching current - 100 mA
  • Max. voltage on output - 24 V
  • Auxiliary voltage at terminal - +5 V (only when the mains supply is present)
  • Users can receive alarms if the values go outside the paramters set. Alarms are indicated through an internal buzzer , 3 LEDs or email alerts via Ethernet port.

Optional accessories and download

AC/DC adapter 230Vac to 5Vdc/2.1A, EU

AC/DC adapter 230Vac to 5Vdc/2.1A, EU

code A1825eu
AC/DC adapter 230Vac to 5Vdc/2.1A, UK

AC/DC adapter 230Vac to 5Vdc/2.1A, UK

code A1825uk
AC/DC adapter 230Vac to 5Vdc/2.1A, US

AC/DC adapter 230Vac to 5Vdc/2.1A, US

code A1825us

300Mbps Wireless N Nano Router

300Mbps Wireless N Nano Router

code TL-WIFI

Wall holder for the Multilogger

Wall holder for the Multilogger

code MLP002
Holder for mounting the Multilogger on magnetic surfaces

Holder for mounting the Multilogger on magnetic surfaces

code MLP003
Wall holder for external probes DIGIL

Wall holder for external probes DIGIL

code SP009

COMET Database

COMET Database

code CDB
COMET Vision - Software for Dataloggers and Multiloggers

COMET Vision - Software for Dataloggers and Multiloggers

code CV
Database Viewer for Comet Database

Database Viewer for Comet Database

code DBV

Magnetic door contact SA-200-A

Magnetic door contact SA-200-A

code SA200A
Magnetic door contact with cable length 10m

Magnetic door contact with cable length 10m

code SA200A-10
Magnetic door contact with cable length 15m

Magnetic door contact with cable length 15m

code SA200A-15
Magnetic door contact with cable length 2m

Magnetic door contact with cable length 2m

code SA200A-2
Magnetic door contact with cable length 20m

Magnetic door contact with cable length 20m

code SA200A-20
Magnetic door contact with cable length 30m

Magnetic door contact with cable length 30m

code SA200A-30
Magnetic door contact with cable length 5m

Magnetic door contact with cable length 5m

code SA200A-5

Adapter for easy connection U/I signals to Multilogger

Adapter for easy connection U/I signals to Multilogger

code MLP006
External audio unit for MSxx and Multilogger

External audio unit for MSxx and Multilogger

code MP026

Software

download-icon Comet Vision Software (2.3.8801.28020 dbapi 21.1)
16.23 MB
download-icon Comet Database One Click Installer (21.1-IIi)
487.31 MB
download-icon Database Viewer 3.20.0.0 dbapi 21.1
24.33 MB

Others

download-icon Terms and Conditions COMET System s.r.o.
133.12 KB

Manuals

download-icon Instruction Manual
2.24 MB
download-icon Instruction Manual - the connection of detectors to the Multilogger M13xx
428.37 KB
download-icon Quick Start Manual
1,009.40 KB
download-icon Declaration of Conformity
58.20 KB
download-icon MIB
575.32 KB

Catalogs

download-icon CATALOG - Multilogger
1.24 MB

Related products

Ethernet Multilogger - thermo hygro meter with 4 MiniDIN inputs

Ethernet Multilogger - thermo hygro meter with 4 MiniDIN inputs

code M1140
Multilogger - Thermometer with 4 Thermocouple inputs and Ethernet port

Multilogger - Thermometer with 4 Thermocouple inputs and Ethernet port

code M1200
Multilogger - thermo hygro meter with 2 MiniDIN and 2 Thermocouple inputs

Multilogger - thermo hygro meter with 2 MiniDIN and 2 Thermocouple inputs

code M1220
Multilogger - universal 4 channel thermometer hygrometer with 2 MiniDIN and 2 Terminals

Multilogger - universal 4 channel thermometer hygrometer with 2 MiniDIN and 2 Terminals

code M1320
Multilogger - thermo-hygro-barometer with 2 MiniDIN and 2 Terminals

Multilogger - thermo-hygro-barometer with 2 MiniDIN and 2 Terminals

code M1321
Ethernet Multilogger - thermo-hygro-CO2 meter with 2 MiniDIN and 2 Terminals

Ethernet Multilogger - thermo-hygro-CO2 meter with 2 MiniDIN and 2 Terminals

code M1322
Multilogger - thermo-hygro-CO2-barometer with 2 MiniDIN and 2 Terminals

Multilogger - thermo-hygro-CO2-barometer with 2 MiniDIN and 2 Terminals

code M1323
Ethernet Multilogger - universal datalogger with 4 inputs

Ethernet Multilogger - universal datalogger with 4 inputs

code M1300

Multilogger Demo